As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.Expiration date: 09/2024.
|
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: no sample was available for evaluation.The submitted pictures show a fractured outer sheath on the delivery system close to the handle.It was known that procedure was conducted sheathless.Based on the provided information, the investigation is closed with a confirmation for outer sheath fracture.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.The instructions for use state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' regarding preparation of the device the instructions for use state that 'prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.' the instructions for use further state: 'prior to stent graft deployment (.), ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible endovascular system failure.' and 'ensure the delivery catheter is straight (.).' regarding accessories the instructions for use states: 'prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended.'; the packaging pictograms indicate an introducer size of 8f and a 0.035" guidewire.H10: d4 (expiration date: 09/2024), g3, h6 (device).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|