• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN GANZ BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN GANZ BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number D97120F5
Device Problem No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Manufacturer Narrative
One swan-ganz catheter was received by our product evaluation laboratory for a full examination.The report of "pacing did not work" was unable to be confirmed.No visible damage or abnormality was observed from catheter body, balloon, windings and returned syringe.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 min.Without leakage.The lot number for this device was not supplied; therefore, further review of the related manufacturing records could not be performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
As reported, during use with this swan-ganz catheter in a patient with bradycardia, the pacing did not work.As per troubleshooting, a new catheter was used that worked perfectly.The decision of changing the catheter was done very fast and this change did not suppose any risk to the patient.There was no allegation of patient injury.The device was available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ BIPOLAR PACING CATHETER
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key15860241
MDR Text Key306453230
Report Number2015691-2022-09481
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K152633
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD97120F5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/01/2022
Initial Date FDA Received11/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-