EDWARDS LIFESCIENCES, PR SWAN GANZ BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY
|
Back to Search Results |
|
Model Number D97120F5 |
Device Problem
No Pacing (3268)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/31/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One swan-ganz catheter was received by our product evaluation laboratory for a full examination.The report of "pacing did not work" was unable to be confirmed.No visible damage or abnormality was observed from catheter body, balloon, windings and returned syringe.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 min.Without leakage.The lot number for this device was not supplied; therefore, further review of the related manufacturing records could not be performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|
Event Description
|
As reported, during use with this swan-ganz catheter in a patient with bradycardia, the pacing did not work.As per troubleshooting, a new catheter was used that worked perfectly.The decision of changing the catheter was done very fast and this change did not suppose any risk to the patient.There was no allegation of patient injury.The device was available for evaluation.
|
|