It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited noise and low out of range pacing impedance measurements on the right ventricular channel.After analysis of the device, it was determined that the leads on this device were reversed in header.The patient is being brought to the clinic for further evaluation.A lead revision was performed, the electrode tips were manipulated, and the issue was resolved.At this time, this device remains in service.No further adverse patient effects were reported.
|