DEPUY SPINE INC EXPEDIUM SPINE SYSTEM DOUBLE-LEAD TAP 6.0MM 30-60MM PLUS OR MINUS 1.33 PERCENT; TAP, BONE
|
Back to Search Results |
|
Model Number 279702600 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that on an unknown date, the tap broke off.There was no patient consequence.No further information is available.This report involves one expedium spine system double-lead tap 6.0mm 30-60mm plus or minus 1.33 percent.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a review of the receiving inspection (ri) for 6mm dual-lead tap was conducted identifying that lot number ng29231 was released in one batch.Batch1: lot qty of (b)(4) units were released on 11 oct 2011 with no discrepancies.Supplier: depuy: ng instruments as a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|