BOSTON SCIENTIFIC CORPORATION GUIDEZILLA II; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 1871
Device Problems Device Damaged Prior to Use (2284); Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2022
Event Type  malfunction  
Event Description
It was reported that device fracture occurred.A guidezilla ii was unpacked and the steel wire braid was exposed.The device was not used in the procedure.No patient complications were reported.
 
Event Description
It was reported that device fracture occurred.A guidezilla ii was unpacked and the steel wire braid was exposed.The device was not used in the procedure.No patient complications were reported.
 
Manufacturer Narrative
The device was returned for analysis.The device was visually and microscopically examined.There was blood present in the shaft of the device.There was partial collar and shaft detachment.There were numerous flat shaft segments to the device of which included the tip.Product analysis confirmed the reported event as the collar and shaft were partially detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA II
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key15869234
MDR Text Key307663037
Report Number2124215-2022-49328
Device Sequence Number1
Product Code DQY
UDI-Device Identifier08714729939450
UDI-Public08714729939450
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K163314
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/04/2023
Device Model Number1871
Device Catalogue Number1871
Device Lot Number0027775289
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight80 KG
-
-