• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problems Gas/Air Leak (2946); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Air Embolism (1697); Chest Pain (1776); Fistula (1862)
Event Date 10/12/2022
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that after a successful cryo ablation procedure, the patient suffered from an atrioesophageal fistula (aef) at the left inferior pulmonary vein (lipv).The patient subsequently underwent surgical repair of the fistula and was subjected to prolonged hospitalization. no further patient complications have been reported as a result of this event.
 
Event Description
It was reported that after a successful cryo ablation procedure, the patient suffered from an atrioesophageal fistula (aef) at the left inferior pulmonary vein (lipv).The patient subsequently underwent surgical repair of the fistula and was subjected to prolonged hospitalization. no further patient complications have been reported as a result of this event.2022-12-06: a few weeks following the successful cryo ablation procedure, the patient presented to the emergency room (er) with chest pain.A computerized tomography (ct) scan revealed air in the left atrium.The patient is recovering but does have some lingering deficits; it was further reported that the patient is stable and has an endotracheal tube and a percutaneous endoscopic gastrostomy.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The files showed 12 applications were performed using an afapro28 balloon catheter with lot 16279.No system notices were triggered on the event date.The console seems to be controlling the pressure and flow within the specified range.In conclusion, the clinical issue (atrioesophogeal fistula, embolism) occurred during the procedure.There is no indication of a relation of the adverse event to the performance and malfunction of the product.The physical product was not returned for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15869384
MDR Text Key304354878
Report Number3002648230-2022-00471
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000418052
UDI-Public00763000418052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0011297785
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2022
Initial Date FDA Received11/28/2022
Supplement Dates Manufacturer Received11/29/2022
Supplement Dates FDA Received12/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AFAPRO28 BALLOON CATHETER
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient SexMale
-
-