• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 M20
Device Problems Operating System Becomes Nonfunctional (2996); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2022
Event Type  malfunction  
Event Description
It has been reported to philips that the system went into reboot sequence immediately.No harm has been reported to philips.To date, no further information was received.We are conservatively reporting this event as the investigation is ongoing.A follow up report will be submitted when further information is received.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue occurred.The philips field service engineer (fse) inspected the system onsite and confirmed that the system went into reboot sequence immediately.Upon further inspection fse found issues with the cat5 cable from ups to m cabinet faulty.The fse replaced the cat5 cable from ups to m cabinet.After replacement of the cable, the system was returned to use in good working order.The codes were updated based on the investigation outcome.The device problem code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15872786
MDR Text Key304514986
Report Number3003768277-2022-01227
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838099258
UDI-Public00884838099258
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200917
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722224
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2022
Initial Date FDA Received11/28/2022
Supplement Dates Manufacturer Received11/09/2023
Supplement Dates FDA Received11/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-