|
Model Number 72203270 |
Device Problems
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/02/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
|
|
Event Description
|
It was reported that during a hip arthroscopy surgery, the bolt of the unkn ilizarov exfix instr device that clamps the traction slider in position, came apart and was not able to tighten the clamp sufficiently.The surgeon used a vice grip to resolve this problem.Once that was fixed, the orange straps that hold the foot into the support were sliding through the clasps allowing the heal to be raised out of the support.Therefore, the procedure could not be performed.The patient was already under anesthesia and had to be woken up.The procedure has been postponed.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Manufacturer Narrative
|
H11: internal complaint reference: (b)(4).H6: code correction.
|
|
Manufacturer Narrative
|
H10: internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Correction in h6 (medical device problem code).
|
|
Manufacturer Narrative
|
H11: this report was inadvertently submitted under manufacturer number ¿1020279¿, the correct manufacturer number is ¿3003604053.¿.
|
|
Event Description
|
It was reported that during a hip arthroscopy surgery, the bolt of the universal hip distractor device that clamps the traction slider in position, came apart and was not able to tighten the clamp sufficiently.The surgeon used a vice grip to resolve this problem.Once that was fixed, the orange straps of the ahtb, active heel traction boot device that holds the foot into the support were sliding through the clasps allowing the heal to be raised out of the support.Therefore, the procedure could not be performed.The patient was already under anesthesia and had to be woken up.The procedure has been postponed.
|
|
Manufacturer Narrative
|
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is worn from use.It is attached to the hip distractor.The orange straps are very worn from use.A functional evaluation of the device found that the orange straps are able to be easily loosened by pulling due to the wear on the straps.The clamps are unable to grip fully into the straps.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states there were no clinical factors found which would have contributed to the event.The patient impact beyond the anesthesia and postponement of the procedure cannot be determined with the information provided.The root cause has been associated with component failure.Factors that could have contributed to the reported event include wear and tear from repeat use.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|