• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number LIFEVEST WCD 4000 SYSTEM
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Insufficient Information (4580)
Event Date 11/08/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt has been completed.The reported problem (connector latch does not secure with monitor) has been confirmed.Upon investigation the electrode belt was unable to complete essential performance testing.The cause for the failure was the trunk cable was damaged and unable to connect to a monitor.The root cause for the damaged connector was excessive force.There was no adverse event that resulted from the damaged belt.
 
Event Description
A us distributor reported an electrode belt connector latch does not secure with monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA
Manufacturer Contact
zoll mfg corporation
121 gamma drive
pittsburgh, PA 15238
MDR Report Key15876560
MDR Text Key305692273
Report Number3008642652-2022-25936
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIFEVEST WCD 4000 SYSTEM
Initial Date Manufacturer Received 11/22/2022
Initial Date FDA Received11/29/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-