• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL574T
Device Problem Unintended Ejection (1234)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2022
Event Type  Injury  
Manufacturer Narrative
Visual investigation: collective report - here we found deformed latches/tips of the slider sheet and clip jams.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Explanation and rationale: there are no hints of a pre-damage or similar.Investigations leads to the assumption that the titanium clip cartridge was engaged not completely or was damaged during inserting.If the cartridge is engaged not completely or was damaged during inserting, there is an impairment of product functionality.This could lead to deformations on the slider sheet, to clip jam and can cause the titanium clip cartridge to jump off.Possibly a damaged clip applicator due to a deformed push rod or something similar could be another cause for a detached titanium clip cartridge.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results, a corrective action and preventive action (capa) has been initiated.
 
Event Description
It was reported that there was an issue with pl574t -challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the clips "jump out" from the cartridge during application.The clips were removed from the patient's body.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15879571
MDR Text Key304460824
Report Number9610612-2022-00355
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52752591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-