• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Material Twisted/Bent (2981)
Patient Problem Hyperglycemia (1905)
Event Date 10/31/2022
Event Type  Injury  
Manufacturer Narrative
Applicator (b)(4) has been returned and investigated.Visual inspection has been performed on the applicator and no issues were observed.It is observed that the applicator was fired correctly.The sharp has been investigated and no issues were observed.The issue is not confirmed.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specifications.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
The caller reported a customer being unable to test using adc device due to bent needle.It was further reported that due to this issue, the customer experienced dizziness and nausea.The caller indicated that due to this issue, the customer experienced an adverse event in which they required treatment of 2 units of insulin provided by a healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15883555
MDR Text Key304501809
Report Number2954323-2022-43569
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2023
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-