Model Number 21-7324-01
Device Problems Obstruction of Flow (2423); Deformation Due to Compressive Stress (2889)
Patient Problem Pain (1994)
Event Date 11/10/2022
Event Type  malfunction  
Manufacturer Narrative
No product was returned.We are unable to confirm the reported complaint.If the product is returned, the manufacturer will reopen this complaint for further investigation.Operator of device is unknown.No information has been provided to date.
 
Event Description
It was reported that there was an incident with a pump with epidural.The fentanyl dose was a100 milliliter bag.The pump displayed a message that the bag was almost empty while there was still a remaining volume of approximately 70 milliliters minimum.A biomedical engineering technician believes the problem is with the tubing and it's caused by green anti-leakage clamp.The tubing pinches here and creates an occlusion that cannot be detected by the device.To solve the problem, you have to massage tubing while pressing anti-flow clamp to undo crushing creating obstruction.It was also indicated that they have another pump with the same problem.There was no patient injury reported but the patient had a lot of pain due to not receiving prescribed dose.No medical or surgical intervention was reported.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updates not required.H10: the device was not returned for investigation, so no visual inspection and functional test were performed.Based on the fact that no product was returned, the investigator was unable to confirm the reported complaint.A dhr (device history review) was performed, and no problems or issues were identified during this dhr review.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available, corrected data: correction: b3: date of event updated to 10-nov-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET WITH FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15888965
MDR Text Key307087239
Report Number3012307300-2022-27620
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586029653
UDI-Public10610586029653
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7324-01
Device Catalogue Number21-7324-24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/14/2022
Initial Date FDA Received11/30/2022
Supplement Dates Manufacturer Received11/30/2022
Supplement Dates FDA Received12/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-