Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having minimally invas ive posterior lumbar interbody fusion.It was reported that, the set screw was tightened down to the rod.When the surgeon went to torque off the set screw, the set screw kept spinning and was not breaking off.Different additional set screws were tried but those were not torqued off either.Eventually, the screw had to be replaced and then the set screw torqued off.It was believed to be the extender tabs were slightly bent causing the set screw to cross thread and causing it to spin instead of torqueing off.It is not able to be confirmed definitively that the screw was the cause of the issue but after further investigation, the bent tabs seem to have been the issue and not the set screw.The surgeon stated he medialized both screws which could have caused the tabs to bend.The pedicle screw came into contact with the patient because it was inserted into the pedicle.The screw was later removed and replaced.The alleged screw was not found, most likely it was thrown away or lost.The surgeon followed ifu/labelling through out the procedure.There were no patient symptoms or complications reported.No additional treatment or surgery performed as a result.No further complications reported regarding the event.
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|