Model Number I400+ |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter occupation: occupation is field service engineer (fse).The fse replaced the affected components.The investigation is ongoing.
|
|
Event Description
|
We received an allegation of a burnt component of a cobas integra 400 plus.The field service engineer (fse) was performing a service repair related to an xyz transfer mechanism failure.When the fse rebooted the system, the xyz printed circuit board smoked and the fse observed a burnt integrated circuit (ic) chip.An evacuation was not necessary and no one was harmed by the smoke or the burnt chip.
|
|
Manufacturer Narrative
|
The field service engineer (fse) replaced the xyz printed circuit board (pcb) as well as a probe and initialized the system.The fse installed the uninterruptible power supply (ups) and removed the failed unit.All checks passed.The service actions resolved the issue.
|
|