MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Interrogate (1332); Communication or Transmission Problem (2896); Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: b3300542, serial#: (b)(4), implanted: (b)(6) 2022, product type: lead.Product id: b3300542m, serial#: (b)(4), implanted: (b)(6) 2022, product type: lead.Product id: b31000, lot#: 082t30721, implanted: (b)(6) 2022, product type: accessory.Product id: b32000, lot#: 082m04122, product type: accessory.Other relevant device(s) are: product id: b3300542, serial/lot #: (b)(4), ubd: 28-jul-2024, udi#: (b)(4).Product id: b3300542m, serial/lot #: (b)(4), ubd: 19-aug-2024, udi#: (b)(4).Product id: b31000, serial/lot #: (b)(4), ubd: 03-nov-2023, udi#: (b)(4).Product id: b32000, serial/lot #: (b)(4), ubd: 10-feb-2024, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that connectivity didn't pass and when the manufacturer representative (rep) tested impedance the electrode 0 was either orange or red.Troubleshooting was not required.The issue was not resolved through troubleshooting.
|
|
Manufacturer Narrative
|
Continuation of d10: product id b3300542, serial# (b)(6), implanted: (b)(6) 2022, product type lead product id b3300542m, serial# (b)(6), implanted: (b)(6) 2022, product type lead product id b31000 lot# 082t30721, implanted: (b)(6) 2022, product type accessory product id b32000 lot# 082m04122, product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, metronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the manufacturer representative (rep) who clarified the report of electrode 0 was red, then orange, then post procedure back to red.The cause of the issue was not determined.Actions taken to resolve the issue included extension ends were removed, cleaned, reinserted and impedances rechecked.It didn't pass connectivity even post procedure, impedances on contact 0 remained red.Impedances assumed to be the issue blocking connectivity test passing.
|
|
Search Alerts/Recalls
|
|
|