On 21st november, 2022 getinge became aware of an issues related to manual trolley.The manual trolley is not registered as a medical device, however upon the situation occurrence it was used with the medical device washer disinfector with the model name 9128e.As it was stated, the loading trolley almost fell on the floor due to a faulty stop pin that is installed on the manual trolley to prevents the loading trolleys from falling off.The malfunctioned part was replaced and the manual trolley was returned to the service in full working condition.There was no injury or damage reported, however we decided to report the issue based on a potential as loading trolley falling to the floor could bring a hazardous situation for the operator and lead to serious injury if the situation was to reoccur.
|