EDWARDS LIFESCIENCES SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number PE075F5
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Manufacturer Narrative
The reported event of balloon inflation difficulties was not able to be confirmed.As received, the balloon latex, distal and proximal windings were completely missing and not returned.Residual adhesive was not observed from the distal electrode and distal side of catheter tube.Adhesive was found on the proximal electrode and catheter body around proximal leadwire but it appeared to be an adhesive for proximal leadwire bonding and not for balloon attachment.No visible damage was found from catheter body.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that the balloon of a swan ganz catheter did not inflate before use.Product was returned and balloon latex, distal and proximal windings were completely missing and not returned.Additional information were obtained from the sales rep after product evaluation was performed.Detachment of the balloon was not reported from the customer although they would likely have noticed if the balloon was detached.Even if balloon inflation issue is observed during balloon test, the balloon is not always verified closely by the customer.Therefore, there is a possibility that the customer does not notice if the balloon was missing initially.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key15894737
MDR Text Key306448082
Report Number2015691-2022-09584
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Model NumberPE075F5
Device Lot Number64481575
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-