• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3095040
Device Problem Chemical Problem (2893)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Dmf# - 13704; trade name gentamicin sulphate; active ingredient(s) gentamicin sulphate; dosage form - powder; strength 1.0g active in our cements.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the smart set ghv cement was used.The cement started to stick to the glove of the surgeon within 4 mins from mixing of the cement.There was no surgical delay.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device [3095040/3746760] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GHV GENTAMICIN 40G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15895768
MDR Text Key307895873
Report Number1818910-2022-24103
Device Sequence Number1
Product Code MBB
Combination Product (y/n)Y
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3095040
Device Lot Number3746760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-