It was reported that on (b)(6) 2022, a 31mm epic mitral valve was selected for an implant in a patient with a history of infective endocarditis.During the procedure, after threading the leaflet, it did not take coaptation and one of the leaflet was bent at the base toward the ventricle.The physician assumed that it was due to dryness, and applied physiological saline, but the leaflet did not return to their original shape.Therefore the procedure was completed by implanting a non-abbott device.There was a 30mins delay but was not significant and the patient remained stable throughout the procedure.No patient consequences were reported.
|
An event of incomplete leaflet coaptation and bent leaflet was reported.A more comprehensive assessment, including functional examination of the valve could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
|