Model Number 93331
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problem Impaired Healing (2378)
Event Date 11/09/2022
Event Type  Injury  
Manufacturer Narrative
This report is submitted on december 2, 2022.
 
Event Description
Per the clinic, the patient experienced skin overgrowth.There was a skin revision and the device was explanted under a general anaesthetic on (b)(6) 2022.
 
Manufacturer Narrative
Correction: the initial mdr submitted on december 2, 2022, was filed inadvertently.No explantation has occurred.Only the abutment was removed.The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Excessive skin thickening and skin growing over the abutment are common complications with baha abutments.Issues can be resolved with clinical case management.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on january 4, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA400 IMPLANT 4MM W ABUTMENT 10MM
Type of Device
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
aini yusof
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key15900037
MDR Text Key304679296
Report Number6000034-2022-03597
Device Sequence Number1
Product Code MAH
UDI-Device Identifier09321502022712
UDI-Public(01)09321502022712(10)COH824569(11)02242016(17)01172021
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/17/2021
Device Model Number93331
Device Catalogue Number93331
Device Lot NumberCOH824569
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/01/2022
Supplement Dates Manufacturer Received12/14/2022
Supplement Dates FDA Received01/03/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/24/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age36 YR
Patient SexMale
-
-