• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EPIK UNI KNEE; BASEPLATE, ONLAY TIBIA, SIZE35 LM/RL EPIK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EPIK UNI KNEE; BASEPLATE, ONLAY TIBIA, SIZE35 LM/RL EPIK Back to Search Results
Model Number 300-01-035
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Loss of Range of Motion (2032)
Event Date 11/16/2022
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as loss of range of motion and some subsidence.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Item: unk-surg.The part/lot number was not reported, therefore; this component could not be linked to a specific device history record (dhr).Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loss of range of motion and some subsidence.There were no findings during this evaluation that indicate that the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - loss of range of motion.X-rays showed some subsidence of the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIK UNI KNEE
Type of Device
BASEPLATE, ONLAY TIBIA, SIZE35 LM/RL EPIK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15901679
MDR Text Key304683803
Report Number1644408-2022-01612
Device Sequence Number1
Product Code HRY
UDI-Device Identifier00888912113281
UDI-Public(01)00888912113281
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022437
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/28/2020
Device Model Number300-01-035
Device Catalogue Number300-01-035
Device Lot Number064B1007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK-SURG LOT UNKNOWN
Patient Outcome(s) Required Intervention;
-
-