• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Electromagnetic Compatibility Problem (2927)
Patient Problems Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 97810, serial#: (b)(4), implanted: (b)(6) 2021, product type: implantable neurostimulator.Other relevant device(s) are: product id: 97810, serial/lot #: (b)(4), ubd: 14-jun-2022, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that the patient said they need their enterra device replaced (due to normal battery depletion).Patient reported that when their interstim system was first implanted, it was interacting with their enterra device and the enterra device started shocking them.Patient stated their managing hcp adjusted stim on their enterra device to turn it down a little bit and patient no longer felt shocking.Patient said typically an enterra device lasts them a year, but because it was able to be decreased it lasted them 6 months longer.In (b)(6) of 2022, the enterra hcp determined the enterra device was depleted.Agent emailed medtronic rep to contact patient with physician listings for replacement gastric stimulator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15911283
MDR Text Key308067365
Report Number3004209178-2022-15886
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00763000286798
UDI-Public00763000286798
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2022
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/29/2022
Initial Date FDA Received12/02/2022
Date Device Manufactured08/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexFemale
-
-