Model Number CD2411-36Q
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  Injury  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported that inappropriate shock was delivered for af.It was noted the rate fell into the vf zone and therefor svt discriminators were not applied.Programming changes were suggested.The episode will be reviewed with the physician.No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15912269
MDR Text Key304776692
Report Number2017865-2022-47582
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507615
UDI-Public05414734507615
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Model NumberCD2411-36Q
Device Catalogue NumberCD2411-36Q
Device Lot NumberA000029087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/15/2022
Initial Date FDA Received12/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; TENDRIL