ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE
|
Back to Search Results |
|
Model Number 1456Q/86 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2022 |
Event Type
malfunction
|
Event Description
|
Related manufacture reference number: 2017865-2022-47615.It was reported that the patient presented prior to generator exchange procedure.Upon interrogation, it was noted that the right ventricular lead exhibited high pacing impedance and high capture threshold.During procedure, it was noted lead was not secured in the header.The reported lead was capped and replaced on 16 nov 2022.During replacement, it was noted that the newly implanted left ventricular could not be implanted as the shape of the lead prevented the lead from advancing.The left ventricular lead was not installed and the procedure was completed with an alternative lead.The patient was in stable condition.
|
|
Event Description
|
Related manufacture reference number: 2017865-2022-48393.Related manufacture reference number: 2017865-2022-47615.It was reported that the patient presented prior to generator exchange procedure.Upon interrogation, it was noted that the right ventricular lead exhibited high pacing impedance and high capture threshold.During procedure, it was noted lead was not secured in the header.The reported lead was explanted and replaced on 16 nov 2022.During replacement, it was noted that the newly implanted left ventricular could not be implanted as the shape of the lead prevented the lead from advancing.The left ventricular lead was not installed and the procedure was completed with an alternative lead.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported event was unable to implant was not confirmed.As received, a complete lead was returned for analysis.Electrical testing, visual inspection, and x-ray did not find any anomalies.
|
|
Search Alerts/Recalls
|
|
|