MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE Back to Search Results
Model Number M00553650
Device Problems Positioning Problem (3009); Activation Failure (3270); Poor Visibility (4072)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2022 that an axios stent and electrocautery enhanced delivery system was to be implanted transgastric to the pancreas to treat a 7.5 cm walled-of-necrosis during an endoscopic ultrasound (eus) procedure performed on (b)(6) 2022.During the procedure, the first flange was deployed; however, the physician had difficulty visualizing the first flange under eus view and they were unable to confirm if the stent first flange expanded.Deployment of the second flange was performed; however, the stent moved out of its position and into the patient's stomach.The stent was removed using a rat tooth grasper and the procedure was not completed.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Type of Device
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15914813
MDR Text Key304791401
Report Number3005099803-2022-07169
Device Sequence Number1
Product Code PCU
UDI-Device Identifier08714729904595
UDI-Public08714729904595
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00553650
Device Catalogue Number5365
Device Lot Number0030289215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient RaceBlack Or African American
-
-