Model Number 2426-0007
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: three photos were submitted for quality investigation.The customer complaint of tubing kinked was partially verified by visual inspection.Evaluation of the photos submitted indicate two kinks.One kink is located on the tubing below the drip chamber, another kink is located on the tubing with the pinch clamp.The claim of the kinked tubing causing the pumps to alarm could not be verified due to no sample being received.A device history record review for model: 2426-0007 lot number: 22079245 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 13jul2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause for the issues seen in the photos provided by the customer could not be determine because the physical samples were not available for further inspection.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event Description
It was reported that the bd alaris¿ lvp 20d 3ss cv experienced kinked tubing.The following information was provided by the initial reporter: sets off the alarms due to the kinking in the material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ LVP 20D 3SS CV
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15920210
MDR Text Key307899534
Report Number9616066-2022-01853
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403227998
UDI-Public10885403227998
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2426-0007
Device Catalogue Number2426-0007
Device Lot Number22079245
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2022
Initial Date FDA Received12/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-