Catalog Number 22443-19
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a query of the complaint handling database for the reported lot revealed no other similar complaints reported from this lot.Based on the information provided, a definitive cause for the difficulty removing the delivery catheter after successful filter deployment could not be determined.It may be possible that the delivery catheter and barewire were kinked or bent preventing the ability to remove the delivery catheter from the barewire; however, without having the product to examine, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a lesion in the left common carotid artery with moderate calcification and heavy tortuosity.The large emboshield nav6 embolic protection system (eps) was advanced to the it's intended location and the filter was deployed; however while removing the delivery catheter the filter was noted to be coming back with the delivery catheter.The physician unsuccessfully attempted 2-3 times to remove the delivery catheter without the filter.Therefore, the filter, delivery catheter and the guide catheter were removed as a single unit.There was no adverse patient effect and no clinically significant delay reported in the procedure.Another emboshield nav6 followed by pre-dilatation with a non-abbott balloon and implantation of an xact stent were used to complete the procedure.No addtional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15921165
MDR Text Key306828945
Report Number2024168-2022-12241
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number22443-19
Device Lot Number2060761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2022
Initial Date FDA Received12/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight75 KG
-
-