The japan customer reported "inconsistent staining.Additionally, field service engineer noted "dripping of liquid." the field service engineer repaired the instrument by replacing the aspiration and dispense check valve, stopcock, syringe, probe, liquid sensor board, and cable for liquid sensor board which resolved this malfunction.Per documentation, customer is able to continue testing.The instrument is fully operational, within specification and ready for the user.Diagnostics were not altered.No direct or indirect patient harm or user harm have been reported.
|