Model Number 00884450123515
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/04/2022
Event Type  Injury  
Manufacturer Narrative
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during insertion of the catheter during a procedure, the catheter broke into two separate pieces.The physician could not remove the pieces and referred the patient to a surgeon to have the pieces removed.
 
Event Description
The account alleges that during insertion of the catheter during a procedure, the catheter broke into two separate pieces.The physician could not remove the pieces and referred the patient to a surgeon to have the pieces removed.
 
Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key15928445
MDR Text Key304954919
Report Number3011642792-2022-00109
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00884450123515
UDI-Public00884450123515
Combination Product (y/n)N
Reporter Country CodeIR
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00884450123515
Device Catalogue Number510038ULT4-T40/B
Device Lot NumberI2133432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-