Model Number 560BCS |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/21/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that prior to use of a bio-console instrument, it was reported that it would not run on battery backup.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that there was no error warning as the instrument just shut off when the hospital power cord was removed.
|
|
Manufacturer Narrative
|
Device evaluation summary: the reported issue that the instrument would not run on battery backup was verified during service.The service technician verified that the instrument shuts off when switched to battery backup.Further investigation proved that battery 1 had failed with a below operating voltage of 10.2 volts where operation voltage is 13.13 volts.The issue was resolved by replacing both batteries and verifying that both voltages were at 13.13 volts.Preventive maintenance was performed per specifications.Note: the instrument was not returned to a medtronic facility but serviced by field service technician.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Conclusion: the complaint was confirmed for the bio-console instrument's depleted batteries.The issue was verified during service and resolved by replacing the batteries.The batteries are expected to perform for up to three years from the date of battery manufacture.A review of complaint and service records associated with this instrument found no other instances of battery replacement within three years.The batteries met their life cycle requirements.There were no patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|