MEDOS INTERNATIONAL SÃ RL CH EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR A4 5.5 X 37-41MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
Model Number 189401404 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/14/2022 |
Event Type
malfunction
|
Event Description
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022 during a posterior spinal fusion for asd, while final tightening the sfx set screw in question, the set screw got stripped.A spare screwdriver and new sfx set screw were used.The surgery was completed successfully without any delay.No further information is available.This report involves one expedium sfx cross connector system connector a4 5.5 x 37-41mm.This is report 1 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: osh, mni, kwp, kwq, mnh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation was performed for the finished device, it was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.Product code: 189401404.Lot number: om9625.The product was released on: 11/08/2017.Qty: (b)(4).The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the sfx,5.5,ti, med, size a4 had one of the pin screws stripped from the stardrive recess.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A dimensional inspection for the sfx,5.5,ti, med, size a4 was unable to be performed due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the sfx,5.5,ti, med, size a4 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|