• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Brain Injury (2219)
Event Date 11/14/2022
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported a perforator (id 261221) did not stop when it was in contact with dura mater causing a defect in dural sealing, possible dural leakage and risk of neurological deficit.The event required medical intervention to seal the dura.
 
Event Description
N/a.
 
Manufacturer Narrative
The perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - visual inspection utilizing unaided eye performed, unit was soiled, and had a destroyed eo label.Organic material between the inner / outer drill.Spring test attempted; unit rotated fine but the spring would not compress from organic material present.Unit was freed by drilling 5 test holes, then passed the spring test and functioned as designed.Functional test performed.Five test holes were drilled to free the inner / outer and allow the spring to compress.Once freed the unit successfully drilled 5 additional holes and was found to function as intended.Complaint could not be verified.Unit was found to meet all acceptance criteria.Therefore the complaint condition could not be confirmed.Root cause - the root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15931882
MDR Text Key304972278
Report Number3014334038-2022-00275
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number261221
Device Catalogue Number261221
Device Lot NumberJ83A41
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/16/2022
Initial Date FDA Received12/07/2022
Supplement Dates Manufacturer Received03/13/2023
Supplement Dates FDA Received03/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-