The customer reports during unspecified procedures using a 26fr.Resection sheath, the tip of the sheath broke inside the patient.The shards fell inside the patient, but the physician was able to retrieve all the shards.After the issue, the physician was able to complete the procedure with a similar item without any further issues.This has occurred on three occasions.Case with patient identifier (b)(6) reports occurrence 1 of 3.Case with patient identifier (b)(6) reports occurrence 2 of 3.Case with patient identifier (b)(6) reports occurrence 3 of 3.
|
This supplemental report was submitted to provide the results of the legal manufacturer¿s (oste) investigation.The lot number for the subject device is unknown, therefore, a review of the device history records could not be performed.Based on the described damage pattern, it is presumed that the damage to the insulation insert was induced thermally and/or mechanically; therefore, it is most likely attributable to wear and tear and/or improper handling by the customer (more specifically the device being subjected to mechanical overload, impact, accidental dropping, etc.).As a general note, cracks on the insulation material are mostly not visible, making visual inspection difficult.Lost fragments of the ceramic insulation insert can be localized and removed using a suitable x-ray procedure or computed tomography.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.To mitigate the injury to the patient and also device damage, the instruction for use provides the following: - properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.- visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches.- visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).- impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.- do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor complaints related to the device and reported phenomenon.
|