(b)(4).Batch # x95923.Additional information was requested and the following was obtained: "how was the device removed from the patient? => no further information will be available.Did the jaws eventually open? => no further information will be available.2nd device: the clip that was stuck onto the jaw, was malformed or scissored clip? => no further information will be available.How was the device removed from the patient? => no further information will be available.Did the jaws eventually open? => no further information will be available." investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no damage in the external components; upon visual inspection, one clip was noted to be jammed.The clip was removed in order to perform functional testing.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 7 conforming clips.In addition, the device locked out as intended.The jammed clip may have caused the device to not fire the clips properly or at all; however, no conclusion could be reached as to what may have caused the clip jamming.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: insert the clip applier through an appropriately-sized trocar.The empty jaws will passively collapse as they are inserted through a 5 mm trocar and reopen when completely through the trocar.Prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|
It was reported by the sales rep that during a laparoscopic cholecystectomy, the jaws did not open after firing.The device was replaced, but the same issue occurred.The trigger was grasped forcibly.After firing, one side of the clip was stuck to the jaw.The device was used on the blood vessels.Another device was used to complete the case.There were no adverse consequences to the patient.
|