• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 3822
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2022
Event Type  malfunction  
Event Description
It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery.A 10x2.25 flextome cutting balloon was selected for use.During procedure, the balloon ruptured upon fifth inflation at 9 atmospheres within 10 seconds.The balloon was removed from the patient's body without any problem using the normal method and the procedure was completed with another of the same device.No complications reported and the patient was in good condition after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME CUTTING BALLOON
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15940087
MDR Text Key307911361
Report Number2124215-2022-50451
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/09/2023
Device Model Number3822
Device Catalogue Number3822
Device Lot Number0025185477
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-