Model Number 26605
Device Problems Positioning Failure (1158); Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/12/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).
 
Event Description
It was reported that stent damage occurred.The 85% target lesion was located in the severely tortuous and moderately calcified internal carotid artery.An 8.0-29 carotid wallstent was advanced for treatment.However, during the procedure slight resistance was encountered, and the stent could not be deployed.The device was removed, and it was found out that the stent was deformed, the distal diameter was shorter than the proximal diameter.The procedure was completed with a different device.No patient complications were reported, and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15941081
MDR Text Key307917977
Report Number2124215-2022-48554
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26605
Device Catalogue Number26605
Device Lot Number0029959750
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2022
Initial Date FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
Patient Weight76 KG
-
-