H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual inspection found cracking on luer lock adapter female, this cosmetic issue could cause leakage or another functional fail test, failure mode is confirmed.The root cause of the reported issue was found to be supplier item fault.Scar submitted to supplier for luer connector this was submitted to supplier on 27-nov-2020, in which supplier was notified about the issue reported for broken component, lot number reported showed that pieces were built with luer connector with lot number raw material 0004081226 with manufacturing date on april 2019, so based on manufacturing date, this action can be assigned to this complaint.Reported lot (raw material) was manufactured before the corrective actions above.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|