Model Number A3059
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: unit received had the index knob replaced due to having both tabs broken off.The disk ratchet and retainer was replaced as well due to wear.All worn components was replaced with new parts.General cleaning and maintenance performed.Root cause: complaint confirmed via inspection of the unit.The unit had worn parts as well as lock movement due to wear on the starburst teeth.Probable root cause is normal wear and tear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event Description
This is 1 of 3 reports linked to mfg report numbers: 3004608878-2022-00262, 3004608878-2022-00263.A facility reported that the swivel lock of the mayfield composite series skull clamp (a3059) moves while in locked position.Rocker arm also moves while in locked position, and starburst teeth are worn down.No information on patient involvement, injury, or surgical delay was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15955556
MDR Text Key308044904
Report Number3004608878-2022-00261
Device Sequence Number1
Product Code HBL
UDI-Device IdentifierM268A30591
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2022
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-