MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Model Number VERSION 1.0
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  malfunction  
Event Description
It was reported that during a recanalization procedure, the device allegedly broke.Reportedly, broken pieces were snared and the device was removed.There was no reported patient injury.
 
Manufacturer Narrative
The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 08/2024).
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation and no images were provided for review.Therefore, the investigation is inconclusive for the reported issue.A clear root cause could not be identified but a helix break represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2024), g3, h6 (method).H11: h6 (result).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a recanalization procedure, the device allegedly broke.Reportedly, broken pieces were snared and the device was removed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15956908
MDR Text Key308108626
Report Number3008439199-2022-00178
Device Sequence Number1
Product Code MCW
UDI-Device Identifier07640142810582
UDI-Public(01)07640142810582
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80237
Device Lot Number220955
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-