|
Lot Number 0030185997 |
Device Problems
Use of Device Problem (1670); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/04/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a cryoablation procedure to treat an atrial fibrillation in the atrium, a polarsheath was selected for use.After the sheath was inserted into the body, the balloon catheter was inserted into the sheath.At that time, back blood was drawn by connecting a syringe to the side port, but air was drawn into the side port.Also, when air was being removed a kink occurred.Sheath was replaced and the issue was resolved.Procedure was able to be completed successfully without any patient complications.Sheath was returned to boston scientific corporation for laboratory analysis.
|
|
Manufacturer Narrative
|
Polarsheath steerable sheath was evaluated by boston scientific corporation.Functional test was performed, and the sheath passed all standard manufacturing testing.The sheath was able to hold pressure while pressurized at 5.5 psi in hemostasis testing.Visual inspection noted there was no kink observed during inspection of the returned sheath.Based on all the available evidence the conclusion is that no problem was detected with the returned device, the reported allegation of air in the device could not be replicated, and no other issues were found with the device.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).The ifu states "do not aspirate via the side port if the sheath lumen is occupied (i.E., by the dilator or components of the cryoablation catheter) as the aspiration may draw air across the sheath valve into the polarsheath.", as investigation found no defect with the device, if any air was seen in the sheath the most likely cause would have been use of the device that was contradictory to the provided instructions.
|
|
Event Description
|
It was reported that during a cryoablation procedure to treat an atrial fibrillation in the atrium, a polarsheath was selected for use.After the sheath was inserted into the body, the balloon catheter was inserted into the sheath.At that time, back blood was drawn by connecting a syringe to the side port, but air was drawn into the side port.Also, when air was being removed a kink occurred.Sheath was replaced and the issue was resolved.Procedure was able to be completed successfully without any patient complications.Sheath was returned to boston scientific corporation for laboratory analysis.
|
|
Search Alerts/Recalls
|
|
|