Model Number CI-1400-01 |
Device Problems
Expulsion (2933); Material Protrusion/Extrusion (2979)
|
Patient Problem
Unspecified Infection (1930)
|
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient reportedly experienced skin flap breakdown and device extrusion.The doctor suspected the infection was due to the recipient's bump on the head months prior.The recipient received medical intervention twice.The recipient's device was explanted.The recipient was not reimplanted.The recipient is doing well.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage in the silicone overmold on the top cover, as well as a severed electrode.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|