SMITH & NEPHEW, INC. JII UNI TIB BASE SZ 6 LM/RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71935198 |
Device Problem
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
Inadequate Osseointegration (2646)
|
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through the review of clinical evidence from post market clinical data collection activities that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
|
Event Description
|
It was reported that, after a left medial unicompartmental knee replacement had been performed on an unspecified date to address osteoarthritis symptoms, the patient experienced aseptic loosening of the tibial baseplate.A revision surgery was performed on an unspecified date to treat this adverse complication.During this procedure, the tibial base plate was explanted and replaced with another smith and nephew implant.According to the data set provided, this event occurred at almost nine (9) months following the index implantation.The outcome of this patient is unknown.No further information is available.
|
|
Search Alerts/Recalls
|
|
|