The catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex that are cleared in the us.The pro code and 510 k number for the rotarex are identified in d2 and g4.The medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was returned for evaluation and a physical investigation was performed for the catheter.The helix was broken at 47 cm distance from the tip of the catheter.Therefor, the investigation is confirmed for thr reported break issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiration date: 07/2024), (method, result, conclusion).Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|