• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJOHUNTLEIGH POLSKA SP. Z O.O. CITADEL PLUS; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJOHUNTLEIGH POLSKA SP. Z O.O. CITADEL PLUS; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number FXX21A4A4AKFBB
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/12/2022
Event Type  malfunction  
Manufacturer Narrative
Process of gathering and analyzing information is ongoing.Additional information will be provided upon conclusion of the investigation.
 
Event Description
It was claimed by the customer staff that the side rail was broken.The device evaluation revealed that the side rail panel was detached.The bed was damaged when a patient was getting up from the bed frame.No injury was claimed.
 
Manufacturer Narrative
It was claimed by the customer staff that the side rail was broken.The device evaluation revealed that the side rail panel was detached.The bed was damaged when a patient was getting up from the bed frame.No injury was claimed.The review of post-market surveillance data and the investigation carried out at the manufacturer site revealed that the main factor which could lead to the side rail detachment might be related to an excessive force applied to the side rail.This is in line with the event description, the side rail was damaged by the patient who shifted all of their weight on the side rail during getting up.The photographic evidence revealed that the side rail panel was mechanically damaged, it was detached from the side rail mechanism (the screws holding the panel were ripped off).The instruction for use for citadel plus bed frame (ifu document number: 831.374_en) includes preventive maintenance procedures for side rails: the operation of the side rails should be checked daily by the device owner.If the malfunction is identified, arjo or approved service agent should be contacted.Ifu includes also safety information: ¿caregiver should always aid patient in exiting the bed.¿ based on the analysis of the complaints, the external excessive force must first compromise the integrity of the safety side prior to breaking it.Arjo device failed to meet its performance specification since the side rail panel was detached from the side rail mechanism.The citadel plus bed was used with the patient when the event occurred and in that way contributed to the event.This complaint is deemed reportable due to the safety side detachment from the bed frame.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CITADEL PLUS
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL  PL-62052
Manufacturer (Section G)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL   PL-62052
Manufacturer Contact
justyna kielbowska
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
883337089
MDR Report Key15968014
MDR Text Key308165388
Report Number3007420694-2022-00206
Device Sequence Number1
Product Code FNL
UDI-Device Identifier05055982756679
UDI-Public(01)05055982756679(11)210608
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFXX21A4A4AKFBB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-