MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMA1Q1 |
Device Problem
Defibrillation/Stimulation Problem (1573)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/24/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of concomitant medical product: product id: 694758.Lead implanted: (b)(6) 2010; medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there was t-wave oversensing (twos) on the right ventricular (rv) lead during ventricular fibrillation (vf) and ventricular tachycardia (vt) non sustained episodes.This lead to an inappropriate anti-tachycardia pacing and inappropriate shock.The lead was reprogrammed and remains in use.No further patient complications have been reported as a result of this event.It was further reported that an internal review of data transmitted from the device indicated that short circuit protection was triggered during high voltage therapy and less than programmed high voltage energy was delivered.The device remains in use.
|
|
Manufacturer Narrative
|
Product event summary the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated that the ventricular tachyarrhythmia therapy energy was low.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|