MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Failure to Interrogate (1332); Communication or Transmission Problem (2896); Electromagnetic Compatibility Problem (2927)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2022 |
Event Type
Injury
|
Event Description
|
Information was received from a manufacturer's representative (rep) regarding an implantable neurostimulator (ins).The reason for call was rep said patient got cardioversion on wednesday, patient was programmed with cardioversion group and it was used for the procedure.After the procedure patient was able to go back to normal programming with the patient's handset.Rep said about 48 hours later, today, patient got another cardioversion procedure.Rep did asked if the cardioversion group was used, but he didn't get a clear answer from patient's spouse.Rep also didn't know if the cardioversion protocol was used, such as if the pads were perpendicular and if healthcare provider (hcp) waited 60 seconds in between each cardioversion sequence.Rep said hcp did it 3 times.Rep said patient can't connect to implant after cardioversion.Rep to see if he can connect with his tablet, ifhe can't then he'll do a reset neurostimulator command with his tablet.Technical services(ts) told rep if connection is not possible then revision is likely nec essary.The caller was not with the patient.No symptoms reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep) reported the cause of the connection issue following cardioversion wasn¿t known.The issue wasn¿t resolved even after the rep tried to login with the patient programmer, clinician tablet, and resetting the nearest simulator and login.Due to this a decision was made to replace the battery on (b)(6) 2022.
|
|
Manufacturer Narrative
|
H3: the implantable neurostimulator (ins)(serial# (b)(6)) was received, and the complaint of no telemetry was confirmed.Destructive analysis of the ins determined that there was abnormal function of an integrated circuit on the hybrid circuit board of the ins.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H6: correction to annex d code.D01 is the applicable annex d code for this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|