• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM RETR; CATHETER, HEMODIALYSIS, IMPLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM RETR; CATHETER, HEMODIALYSIS, IMPLA Back to Search Results
Model Number IPN920475
Device Problems Deformation Due to Compressive Stress (2889); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2022
Event Type  malfunction  
Event Description
It was reported that "during the insertion of the dilator, they encountered resistance to introduction.As they pulled out the entire device, they noticed that the guide-wire was bent in several point." no patient harm was reported.A new guidewire was used.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that "during the insertion of the dilator, they encountered resistance to introduction.As they pulled out the entire device, they noticed that the guide-wire was bent in several point." no patient harm was reported.A new guidewire was used.The patient's condition is reported as fine.
 
Manufacturer Narrative
Qn#(b)(4).The customer returned one guide wire and all four of the dilator types provided within the kit (16fr safe sheath d-pro, tunneler dilator, 12fr dilator, and 14fr dilator).Signs of use in the form of biological material was observed on the safe sheath d-pro dilator.Visual analysis revealed that the guide wire was severely kink bent in several locations.Microscopic examination confirmed the damage and revealed that the distal and proximal welds were secure and intact.No defects or anomalies were observed with any of the returned dilators.The major kinks in the guide wire measured 271mm, 373mm, 483, 525mm, and 622mm from the proximal weld.The guide wire total length measured approximately 100.5cm, which is within the specification limits of 98.75mm-101.25mm per the guide wire product drawing.The guide wire outer diameter measured.0370", which is within the specification limits of.0370"-.0385" per the guide wire product drawing.The 14fr dilator outer diameter measured.185", which equals the nominal value of.185" per the dilator product drawing.The 14fr dilator inner diameter at the distal tip measured.041", which is within the specification limits of.040"-.042" per the dilator product drawing.The 12fr dilator outer diameter measured.158", which equals the nominal value of.158" per the dilator product drawing.The 12fr dilator inner diameter at the distal tip measured.041", which is within the specification limits of.040"-.042" per the dilator product drawing.The dilator from the safe sheath d-pro assembly outer diameter.212", which is within the specification limits of.209"-.212" per the safe sheath d-pro product drawing.The dilator inner diameter at the distal tip measured.038", which is within the specification limits of.038"-.039" per the safe sheath d-pro product drawing.The tunneling dilator outer diameter measured.2745", which is within the specification limits of.2720"-.2780" per the tunneling dilator product drawing.The dilator inner diameter at the distal tip measured.130" , which equals the nominal value of.130" per the tunneling dilator product drawing.The guide wire was inserted through the 14fr, 12fr, and 16fr safe sheath d-pro dilators.Minor resistance was encountered at the locations of the kinking; however, the guide wire was able to pass through all dilators.No resistance was encountered in areas not affected by kinking.Performed per ifu statement, "carefully place 12 fr.Tissue dilator onto guidewire and advance to appropriate depth.A twisting motion may assist in advancing through tissue.Repeat this procedure using 14 fr tissue dilator.Thread dilator/sheath assembly over guidewire".The tunneling dilator is not meant to pass over the guide wire, but instead the catheter.A manual tug test confirmed that the distal and proximal welds were secure and intact.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "do not use excessive force when introducing guidewire or tissue dilator as this can lead to vessel perforation, bleeding, or component damage".The ifu also states, "do not use excessive force when introducing guidewire or tissue dilator as this can lead to vessel perforation, bleeding, or component damage".The report of a kinked guide wire was confirmed through complaint investigation of the returned sample.Visual analysis revealed that the guide wire was kinked in multiple locations across the body.Functional evaluation of the dilators revealed that the unkinked portions of the guide wire were able to pass as expected, however, slight resistance was met while advancing the kinked portions.The guide wire and the dilators met all relevant dimensional requirements, and a device history record review was performed with no relevant findings.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-L 15 FR X 23 CM RETR
Type of Device
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15973524
MDR Text Key306155967
Report Number9680794-2022-00775
Device Sequence Number1
Product Code MSD
UDI-Device Identifier30801902195136
UDI-Public10801902195132
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K141051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN920475
Device Catalogue NumberCS-15232-VFE
Device Lot Number13F22F0595
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received12/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-