Catalog Number 88710100 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that it was not functioning at normal speed.The event timing was during surgery.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.No additional information is available.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
|
|
Event Description
|
There is no additional event information.
|
|
Manufacturer Narrative
|
This incident has been recorded under (b)(4).Review of the most recent repair record determined the motor was inconsistent.The motor was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|