BAXTER HEALTHCARE CORPORATION CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 2C8541 |
Device Problems
Disconnection (1171); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that three (3) clearlink system continu-flo solution sets were cut and missing the luer-lock adapter.This issue was further described as, ¿in the right-hand side, the tubing is cut and there is no luer-lock adaptor¿.The issue was identified during use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation. visual inspection was performed to the photograph using the naked eye which revealed that the male luer was separated from the tubing.The reported condition was verified. by the nature of the sample, no additional tests were performed. the cause of the condition could not be determined. a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|