Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Occupation is patient/consumer.The meter was requested for investigation.Replacement product was sent.
|
|
Event Description
|
There was an allegation of a display issue with a coaguchek xs meter.The reporter stated their meter's display "goes away and comes back".A display check was performed and the display flickered repeatedly.All segments of the result field are visible when the display is visible.There was no misinterpretation of the results.
|
|
Manufacturer Narrative
|
The customer's meter was returned for investigation.Observed that the meter powered on with fresh retention batteries.A full display check was performed and revealed no missing or faded segments.The display was observed to have been functioning properly.No flickering display issues were observed.The customer's returned meter was tested with retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.8 inr, qc 2: 5.9 inr, qc 3: 5.9 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The customer's allegation could not be reproduced.The returned meter performed according to specifications.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|