Catalog Number 03666794001
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2022
Event Type  malfunction  
Manufacturer Narrative
Occupation is patient/consumer.The meter was requested for investigation.Replacement product was sent.
 
Event Description
There was an allegation of a display issue with a coaguchek xs meter.The reporter stated their meter's display "goes away and comes back".A display check was performed and the display flickered repeatedly.All segments of the result field are visible when the display is visible.There was no misinterpretation of the results.
 
Manufacturer Narrative
The customer's meter was returned for investigation.Observed that the meter powered on with fresh retention batteries.A full display check was performed and revealed no missing or faded segments.The display was observed to have been functioning properly.No flickering display issues were observed.The customer's returned meter was tested with retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.8 inr, qc 2: 5.9 inr, qc 3: 5.9 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The customer's allegation could not be reproduced.The returned meter performed according to specifications.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15982429
MDR Text Key305682000
Report Number1823260-2022-04018
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/23/2022
Initial Date FDA Received12/14/2022
Supplement Dates Manufacturer Received12/22/2022
Supplement Dates FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-